Peristaltic pump MP-4000・4001 series

HOME > Peristaltic pump MP-4000・4001 series

Models MP-4000・4001

• A high-end, top-of-the-line peristaltic pump

• Equipped with flow rate conversion display function, LCD digital display

 

• Supports programmed operation and integration with external devices

♦ The adoption of a DC servo motor enables a motor speed variation ratio of 3000:1. This allows a wide discharge flow rate range using a single tube.
♦ User-friendly LCD display with multi-function indicators and dial controls.
♦ The liquid flow direction (forward/reverse) is shown on the LCD display and can be easily switched using the Fwd./Rev. key.

 

• Intuitive and user-friendly LCD display

 

• External input/output・Contact input terminals

Product name Specification Cat. No.
①Torque / rotation signal output cord 2.5m  120890
External signal input cord set

②External signal input cord +

③Contact signal input cord

One each

3m

2.5m

274420
③Contact signal input cord 2.5m 260640

 

• Two types of tubing holders are available

Adjustable screw type 

The tubing holder is adjusted using a conventional screw. It can be adjusted to a constant discharge pressure regardless of the type or thickness of the tubing.

 

 

 

Spring type 

The tubing holder is pressed with an internal spring mechanism. No adjustment is needed for
tightening, as the tubing is consistently pressed with a fixed force. This reduces flow rate errors
during tubing replacement and minimizes tubing deterioration due to over-tightening.
* The maximum discharge pressure varies depending on the type of tubing.

 

 

 

• One-touch installation of the integrated roller set

The integrated roller set can be smoothly installed by simply turning the switch ON.

 

 

 

• Composition of the roller section

Integrated roller set Cat. No. 252570
Main roller Cat. No. 252550
Roller (6 pcs) Cat. No.252560
Clip (6 pcs)・plate (1 pc) set Cat. No.196150

 

 

• Introduction video of EYELA peristaltic pumps

• Specifications

Model MP-4000 MP-4001
Cat. No. 251400  251410
Tubing holder method Adjustable screw type  Spring type
Flow rate range 0.06 to 1450mL/h x 2
Flow rate accuracy Within ± 2% (repeatability within ± 1%)
Max. discharge pressure 196kPa (2kg/cm2)  98kPa (1kg/cm2)
Max. suction pressure  9.8kPa (0.1kg/cm2)
Viscosity・liquid temperature Max. 2Pa・s (2000c)・0 to 60℃ (no freezing)
Safety features Rotation error detection alarm (overheating, overcurrent, overvoltage, undervoltage, and power failure)
Number of tubes 2 pcs
Number of tubing cases Single stage (can be expanded up to double stages)
Compatible tubing material Silicone, Tygon®, Versilon (Fluran®), PharMed®
Tubing diameter 3 to 6.1mm OD  3.2 to 6.1mm OD
Number of tubing cases Single stage (can be expanded up to double stages)
Ambient temperature range 0 to 40℃ (no condensation)
Outer dimensions (mm) 137(166W x 130(153)D x 177(190)H
Weight Approx. 1.9kg
Power input・supply voltage 1A, 100VA・100VAC 50/60Hz

◊ Tubing is sold separately. Flow rate, discharge pressure, and other performance may vary slightly depending on the tubing used.
◊ When the viscosity of the liquid exceeds 0.1Pa·s (100cP), the discharge volume may decrease.
◊ The dimensions in parentheses ( ) include protrusions.

 

• Options

• Tubing case compatibility

Tubing holder method Model Cat. No.
Adjustable screw type  TCS-0  251220
Spring type TCS-1  251260
Adjustable screw type  TCS-100  251230
Spring type TCS-101 251270

 

 

• Adapter set tubing

Tubing type Silicone, PharMed®, Tygon®, Versilon (Fluran®) By using the adapter set tube, only the worn sections of the tube can be replaced during operation. The tube and tubing case can be connected with a one-touch fitting.
Compatible size (ID x OD) 1.15 x 3.2mm, 2.15 x 4.2mm,
3.15 x 5.2mm, 3.7 x 6.1mm

◊ The adapter set tubing comes as a single set (one channel).

◊ By using the adapter set tubing, only the replaceable adapter set tubing can be exchanged as needed. The tubing and adapter set tubing can be connected easily with a one-touch fitting.

◊ Ensure that the tubing case methods are the same for the upper and lower stages. The tubing case fixing bolts for the double stage and the joint shaft for the lower tubing case are included with the lower stage tubing case.

 

• Data

• Flow rate characteristics

Conditions: Silicone tubing, deionized water, liquid temperature:  approx. 18℃, room temperature: 20℃, back pressure: 0

• Relationship between input voltage and rotation speed