Programmable Low Temperature Precision Water Bath NCB Series

HOME > Programmable Low Temperature Precision Water Bath NCB Series

NCB-3100・3200・3300

• Programmable functions for versatile applications

 

♦ Wide Temperature Range Control: Ensures precise regulation within ±0.03°C across a broad range.

♦ PC Connectivity: Includes measurement and control software for processing temperature trend data when connected to a PC.

♦ Precision Temperature Programming: Supports accurate temperature ramping and cooling with up to 7 user-defined patterns (each with up to 10 segments) and enables pattern linking.

♦ User-Defined Program Options: Allows selection of post-program actions: stop, maintain final temperature, or return to initial settings.

 

• Intuitive control panel

Program parameters can be easily set using the numeric keypad.

 

• Programmable operation modes

User Programs
Type A (3 patterns): Stops operation after the program ends. Type B (4 patterns): Activates standby mode to maintain the final temperature after the program ends. In Type B mode, the standby function allows use of the SKIP/END feature to reset the system to its initial temperature and maintain that condition. After loading a new sample, simply press the key to restart the program, enabling seamless repeat operations.

 

• PC interface and external circulation functionality

The NCB series features an RS-232C interface, allowing temperature control via an external host PC.A stop valve nozzle is provided for external circulation. Supports optional external temperature sensors to control target device temperatures, ensuring stable and precise operation in connected systems.

 

• Remote operation and data processing via dedicated software

♦ Control software ESMon2018
Temperature can be remotely controlled via RS-232C communication, and trend data can be processed using the software.
Note: Only supported in normal mode. Program mode is not supported.

Operating system Microsoft Windows® 10
CPU Pentium or higher (Pentium III 800MHz or higher recommended)
Hard disk A minimum of 15MB of free disk space is required. Additional storage capacity may be necessary for data recording.
Memory Requirements vary depending on the operating system, but 128 MB or more is recommended.
Communication interface  A serial interface (COM1, etc.) is required. If you are using a computer that does not have a serial interface, please use a commercially available “USB-serial conversion cable.”

 

 

• Specifications

 

Product name Programmable Low Temperature Precision Water Bath
Model NCB-3110 NCB-3200 NCB-3310
Cat. No. 283020 112460 283030
Circulation System Closed-loop circulation system 
Temperature control
range
-20 to 80°C -30 to 80°C -20 to 80°C
Temperature control

accuracy

±0.03°C
External circulation
capacity
Max. discharge 6.7L/min (50Hz, water), Max. lifting height 2.9m (50Hz, water)

Temperature slope

1.3°C/min or less (heating),
1.1℃/min or less (40→0°C),
0.5℃/min or less (5→-15°C),
0.4℃/min or less (-15→-20°C)
0.9°C/min or less (heating), 
0.3°C/min or less (cooling)
0.5°C/min or less (heating), 
0.2°C/min or less (cooling)
Cooling capacity at 20°C  370W (318kcal/h) 480W (412kcal/h) 610W (524kcal/h)
Cooling capacity at -10°C  220W (189kcal/h) 350W (301kcal/h) 420W (361kcal/h)
Temperature control PID control with SSR output (JPt 100Ω platinum resistance thermometer)
Temperature setting・display Numeric keypad input; minimum setting unit: 0.01°C; digital display
Program function 1.Fixed value operation (Normal mode) 2.Auto start 3.Auto stop (1min to 99hr59min)
4. User programs: 7 patterns (max. 10 segments/pattern, 1min to 99hr59mim/segment)
Safety features Self-diagnostics (low water level, sensor error, SSR failure, overheating, heater disconnection, cooling failure, power failure, program slope failure), Earth leakage/overcurrent breaker, Refrigerator overload relay
External I/O RS-232C communication port; External temperature sensor terminal
Temperature sensor JPt 100Ω platinum resistance thermometer 
Heater 1.2kW (stainless steel)
Bath stirring method Jet stream circulation
Bath material Stainless steel
Inner bath dimensions
(mm)
250W x 295D x 180H 320W x 405D x 180H 430W x 517D x 180H
Effective bath

dimensions (mm)

215W x 135D x 160H 280W x 240D x 160H 400W x 354D x 160H
Bath capacity 13L 23L 41L
External
circulation nozzle
10.5mm OD discharge port (Rc1/4, with stop valve) and return port (Rc1/4); close system
Cooling coil Stainless steel
Refrigerator / Refrigerant 450W air-cooled (R452A) 400W air-cooled (R407C) 400W air-cooled (R452A)
Ambient operating
temperature
5 to 30℃
Power input 19.5A, 1.95kVA 20A, 2kVA 20A, 2kVA
Power supply 100VAC,  50/60Hz
External dimensions

 (mm)

620(635)W x 390(440)D x 464H 393W x 511D x 793H 513(528)W x 632(672)D x 898H
Weight Approx. 46㎏ Approx. 48㎏ Approx. 59㎏

◊ Performance values are based on measurements at 20°C ambient temperature, 50Hz rated power supply, using water or ethanol as the circulating fluid, and under no-load conditions.

◊ For settings of 10°C or below, use a low-temperature heat transfer medium.

◊ Dimensions in parentheses ( ) include protrusions.

 

• Options

• Insulated hose set

9 mm ID tubing (temperature range: –30 to 80°C)

Length Cat. No. 
1m 112690
2m 112700
3m 174420

 

• Other options

Product name Model Cat. No. Description
Communication cable 167370 Cable for measurement and control software
External temperature sensor

STP-200 113620 For controlling bath temperature based on the temperature at the external circulation point
Clamp set 164500 Used for securing flasks or other vessels immersed in the bath
Standard: Tapered female thread Rc3/8

 

• Data

• Heating and cooling curves

♦ NCB-3110

Conditions: Liquid volume 18L, ambient temperature 20°C, 50Hz, data recorded after the refrigerator warm-up

 

♦ NCB-3200

Conditions: Liquid volume 18L, ambient temperature 20°C, 50Hz, data recorded after the refrigerator warm-up

 

♦ NCB-3310