• A wide flow range, excellent repeatability with digital RPM display
• Achieves a wide flow rate range from 0.06 to 1450 mL/h
♦ The adoption of a DC servo motor enables a motor speed variation ratio of 3000:1. This allows a wide discharge flow rate range using a single tube.
♦ The bright and easy-to-read 7-segment LED display shows the rotation speed. The control knob allows precise speed setting, ensuring highly reproducible flow rates.
• Three types of tubing holders are available
▮ Adjustable screw type
The tubing holder is adjusted using a conventional screw. It can be adjusted to a constant discharge pressure regardless of the type or thickness of the tubing.
▮ Spring type
The tubing holder is pressed with an internal spring mechanism. No adjustment is needed for
tightening, as the tubing is consistently pressed with a fixed force. This reduces flow rate errors
during tubing replacement and minimizes tubing deterioration due to over-tightening.
* The maximum discharge pressure varies depending on the type of tubing.
▮ Fixed-width groove type
The machining tolerance of the tubing case is specified, so by combining it with our genuine tubing, which have specified tolerances for the tubing thickness, there is almost no variation in flow rate or fluctuation in discharge pressure when replacing the tubing.
• One-touch installation of the integrated roller set
The integrated roller set can be smoothly installed by simply turning the switch ON.
• Composition of the roller section
Integrated roller set | Cat. No. 252570 |
Main roller | Cat. No. 252550 |
Roller (6 pcs) | Cat. No.252560 |
Clip (6 pcs)・plate (1 pc) set | Cat. No.196150 |
• Introduction video of EYELA peristaltic pumps
• Specifications
Model | MP-3000 | MP-3001 | MP-3100 | MP-3101 | MP-3000A | MP-3000B |
Cat. No. | 251360 | 251370 | 252660 | 252670 | 251380 | 251390 |
Tubing holder method | Adjustable screw type | Spring type | Adjustable screw type | Spring type | Fixed-width groove type | |
Flow rate range | 0.06 to 1450mL/h x 2 | 0.06 to 1450mL/h x 4 | 0.06 to 180mL/h | 0.5 to 1450mL/h | ||
Flow rate accuracy | Within ± 2% (repeatability within ± 1%) | |||||
Max. discharge pressure | 196kPa (2kg/cm2) | 98kPa (1kg/cm2) | 196kPa (2kg/cm2) | 98kPa (1kg/cm2) | 147kPa (1.5kg/cm2) | |
Max. suction pressure | 9.8kPa(0.1kg/cm2) |
|||||
Viscosity | Max.2Pa・s(2000cP) |
|||||
Liquid temperature | 0 to 60℃(no freezing) |
|||||
Rotation range | Digital display 00.00 to 0.00rpm |
|||||
Safety features | Rotation error detection alarm, Thermal protector |
|||||
Tube case material | PET resin |
|||||
Number of tubes | 2 pcs | 4 pcs | 1 pc | |||
Number of tubing cases | Single stage (can be expanded up to double stages) | Double stages | Single stage | |||
Ambient temperature range | 0 to 40℃ (no condensation) | |||||
Outer dimensions (mm) | 137(166)W x 130(148)D x 177(190)H | 137(166)W x 130(148)D x 206(219)H | 137W x 130(148)D x 178(191)H | |||
Weight | Approx. 1.8kg | Approx. 2.2kg | Approx. 2kg | |||
Power input・supply voltage | 1A, 100VA・100VAC 50/60Hz |
◊ Tubing is sold separately. Flow rate, discharge pressure, and other performance may vary slightly depending on the tubing used.
◊ When the viscosity of the liquid exceeds 0.1Pa·s (100cP), the discharge volume may decrease.
◊ The dimensions in parentheses ( ) include protrusions.
• Options
• Adapter set tubing
Tubing type | Silicone, PharMed®, Tygon®, Versilon (Fluran®) | By using the adapter set tube, only the worn sections of the tube can be replaced during operation. The tube and tubing case can be connected with a one-touch fitting. |
Compatible size (ID x OD) | 1.15 x 3.2mm, 2.15 x 4.2mm, 3.15 x 5.2mm, 3.7 x 6.1mm |
◊ The adapter set tubing comes as a single set (one channel).
• Tubing case compatibility
Tubing case set | Tubing holder method | Model | Cat. No. | |
① | Change to MP-3000
For MP-2000・3000・4000 |
Adjustable screw type |
TCS-0 | 251220 |
② | Change to MP-3000A
For MP-2000A・3000A |
Fixed-width groove type A | TCS-A | 251240 |
③ | Change to MP-3000B
For MP-2000B・3000B |
Fixed-width groove type B | TCS-B | 251250 |
④ | Change to MP-3001
For MP-2001・3001・4001 |
Spring type | TCS-1 | 251260 |
⑤ | Change to MP-3100
For MP-2000・3000 |
Adjustable screw type | TCS-100 |
251230 |
⑥ | Change to MP-3101
For MP-2101・3101 |
Spring type | TCS-101 | 251270 |
◊ Please ensure that the tubing case types are the same for the upper and lower stages in case of adjustable screw model.
The tubing case fixing bolts for the double stage and the joint shaft for the lower tubing case are included with the lower stage tubing case.
• Data
• Flow rate characteristics
Conditions: Silicone tubing, deionized water, liquid temperature: approx. 18℃, room temperature: 20℃, back pressure: 0