Vacuum controller NVC-3000・3000L

HOME > Vacuum controller NVC-3000・3000L

NVC-3000・3000L

• This is a vacuum controller for rotary evaporators

• The communication function enables control of system peripheral devices

♦Equipped with a 4.3-inch color TFT LCD, it provides a comprehensive system data list display for checking operating conditions (settings and measurements) at a glance. It also features a graph screen to monitor control progress and a standard screen to check pressure and operating status, offering clear visibility tailored to usage conditions.

♦In the Auto mode (Automatic control), you can automatically operate the system by simply pressing the start key, starting with the initial boiling point detection of the solvent and setting the appropriate vacuum level. Furthermore, by using the optional vapor temperature sensor, automatic operation of mixed solvents, which was previously difficult, is also possible.

♦In the Manual mode (Constant value control), the pressure setting can be adjusted at any time, allowing you to modify the pressure as needed while monitoring the concentration state during operation.

♦The Program mode (Ramp control) has 55 programs registered for typical solvents, including environmentally regulated solvents. Additionally, up to 10 custom programs can be registered as needed.

♦The Step Program mode functions as a pressure controller with a step program feature. Each program can include up to 99 steps, allowing combinations of pressure reduction, pressure increase, constant pressure control, decompression, and atmospheric release. A total of five programs can be stored (step program control).

 

▶▶ Introduction of vacuum control modes and features of NVC-3000

 

▶▶ Comparison of rotary evaporator system with and without NVC-3000

 

• Control modes

♦ Manual mode (Constant value control)

♦ Auto mode (Automatic control)

♦ Program mode (Ramp control)

♦ Step Program mode

 

• Control of evaporator system by vacuum controller NVC-3000

By combining the vacuum controller NVC-3000 model with system peripheral devices (equipped with communication functions), it is possible to build an integrated system where each device works in coordination. The vacuum controller allows for control of “system start/stop,” “settings changes (NVC-3000 model),” “measurement value checks,” and “emergency stop during alarm occurrence.”

♦ Contributing to research efficiency
By simply pressing the Run/Stop button on the control panel, the operation of the system peripherals can be started or stopped.

♦ Improved safety during research
If an alarm occurs in a peripheral device, the entire system automatically stops.

 

 

 

 

 

 

 

Monitoring of operating conditions

Monitoring of system peripherals and progress can be performed by connecting to a PC.

You can also check temperature, pressure, and rotation speed, and start/stop operation of the system from the PC.
(Installation of dedicated software is required.)

 

 

♦ Download the dedicated software for NVC-3000 from the following link.

setu p_NVC3000Monitor.zip (v113_18/11/09)

Unzip the downloaded compressed file at an appropriate location. A setup disk folder will be created and several files will be extracted into it. Please refer to the “Installation Instructions.pdf” in the extracted “Setup Disk” and install the software.

OS: Windows XP (SP1) to Windows 10 computers supported The software version must be Ver4.00 or later. Please check the startup screen.

 

♦ Connection flow with system peripheral devices equipped with communication capabilities

01 Rotary evaporator
N-1300E・V・S
N-1210BV・S・T
04 Vacuum controller
NVC-3000
02 Water bath・Oil bath
SB-1300, SB-1350
OHB-1100・OSB-2200, etc.
05 Inverter diaphragm vacuum pump
NVP-1000V・2000V・2100V
DTC-31DC
03 Low temperature circulator
CCA-1123A
CCA-1112A
CA-1117A
CA-1116A
CA-1330
06 Diaphragm vacuum pump
NVP-1000・2000・2100, etc.

 

• Specification

Product Vacuum Controller
Model NVC-3000
Cat. No. 269370
Vacuum setting range 0 to 760mmHg (Torr) 0 to 1013hPa (mbar)
Vacuum measuring range 0 to 800mmHg (Torr) 0 to 1066hPa (mbar)
Vapor temperature measuring range 0 to 150℃ (when vapor temperature sensor is connected)
Setting・control method Dial encoder, key input・solenoid valve ON-OFF or variable output voltage
Control program Fixed value・Solvent program・Auto・Step program
Control Items Pressure・Temperature (Cooling water・ Bath・Vapor)・Auto jack・Rotation ON-OFF・Voltage output・Power ON-OFF
Display Color TFT LCD display (4.3 inches) 
Display items Operating mode, Operating status, Measured pressure, Set pressure, Vapor temperature (when optional sensor is connected), Operating time, Bath temperature, Cooling water temperature, Rotary evaporator rotation speed (when communication cable is connected)
External input・output Low temperature circulator・Bath・Rotary evaporator common communication terminal, NVC-NVC communication terminal, Inverter pump communication terminal, Pump control box communication terminal, Vapor temperature sensor terminal, Micro USB (for PC communication)
Ambient temperature range 5 to 35℃
Outer dimensions (mm)・weight 138.7W x 50 (64.5) D x 114.2H・Approx. 670g
Power input ・supply voltage 0.1A, 10VA・100 to 240VAC  50/60Hz
Power supply cord Can be changed with AC adapter

◊ Dimensions in parentheses ( ) include protrusions.

◊ Diaphragm pump control box PBX, control solenoid valve CV, and NVC-PBX connection cord are required when connecting to a diaphragm type vacuum pump.

◊ For inverter diaphragm vacuum pumps, the NVC-NVPV connection cord is required.

 

• Options

• Vacuum control system configuration

• Cables, branch boxes, control boxes

No. Product Length Model Cat. No.  Remark
When connected to inverter diaphragm vacuum pump NVC-NVPV connection cord 0.5m NNV-0.5M  269410 For NVP-V
2m NNV-2M  269420 For NVP-V
When connected to diaphragm vacuum pump Diaphragm pump
control box
PBX 269380 Control box for solenoid valves and vacuum pumps
NVC-PBX connection cord NP-0.5M  269390 For connecting NVC-3000 and PBX
NP-2M  269400 For connecting NVC-3000 and PBX
Control solenoid valve CV-11 269430 10mm OD hose nozzle
CV-12  269440 13mm OD hose nozzle
When connected to system components other than diaphragm vacuum pumps NVC communication branch box F-BOX  269620 Branch box to connect system peripherals
NVC communication cable COM-0.5M  269450 For communication with evaporators, cooling water circulation devices, water baths and oil baths compatible with the NVC-3000 model
COM-1M  269460
COM-2M  269470
When two vacuum controllers NVC-3000 are connected Vacuum control communication cable A 2m VC-2M  245100
4m VC-4M  246690

 

• Vacuum controller mounting boards

Applicable models Model Cat. No.
For inverter diaphragm vacuum pumps Direct mounting board to NVP-V series NVC3000-NVP-V 270100
Mounting board for evaporator N-1300 NVC3000-N1300 ※ 269480
Mounting board for evaporator N-1210B NVC3000-N1210 ※ 269500
Mounting board for evaporator N-2100 NVC3000-N2100 ※ 269520
Stand base 2 271490
For diaphragm vacuum pumps Mounting board for NVP series・PBX NVC・PBX-NVP 270090
Mounting board for EVP series・PBX NVC・PBX-EVP 270110
Mounting board for evaporator N-1300・PBX NVC・PBX-N1300 269490
Mounting board for evaporator N-1210B・PBX NVC・PBX-N1210 269510
Mounting board for evaporator N-2110・PBX NVC・PBX-NVP 269530
Mounting board for evaporator MF-1000・PBX NVC・PBX-MF1000 270130

※It can also be used when mounting the vacuum controllers of the DPE-1150/2150 models and DPE-1250B/C/D/E models onto the rotary evaporator N-1300/1210B/2110 series.

 

• Vapor temperature sensor set for rotary evaporators

Model Applicable models Cat. No.
TS-S For N-1300S, N-1210BS 211770
TS-V For N-1300E・1300V・1210BV・1210BT 211780

 

• NVC3000-PBX set stand

This is a dedicated stand for mounting the vacuum controller NVC-3000, diaphragm pump control unit PBX, and NVC control solenoid valves CV-11 and CV-12.

Cat. No. 275090

◊ Vacuum controller NVC-3000, diaphragm pump control unit PBX, and solenoid valve for NVC control CV-11 are sold separately.

 

 

• NVC-3000L for medium・large-size rotary evaporators

 

♦ A control solenoid valve is included with NVC-3000L for medium・large-size rotary evaporators and vacuum pumps.

♦ The pressure setting can be adjusted at any time, allowing pressure changes during operation while monitoring the concentration status (Manual control).

Model: NVC-3000L

Cat. No. 279410