Trouble shooting
HOME > Trouble shooting > Low Temperature Circulator > CA-1115-CE
HOME > Trouble shooting > Low Temperature Circulator > CA-1115-CE
| Symptom | Causes |
|---|---|
| Electric leakage hazard occurs. | Stop operation and contact your local dealer or closest customer service center. |
| Excess current flows. | Stop operation and contact your local dealer or closest customer service center. |
| Symptom | Causes |
|---|---|
| Mains connector is not plugged into outlet or not plugged into outlet completely. | Plug main connector after turning off residual current device and power switch. |
| Power is not supplied. | Turn on the breaker of distribution board. |
| Residual current device is turned off. | Turn the device on. |
| Residual current device is impaired. | Stop operation and contact your local dealer or closest customer service center. |
| Power switch is impaired. | Stop operation and contact your local dealer or closest customer service center. |
| Temperature controller is impaired. | Stop operation and contact your local dealer or closest customer service center. |
| Symptom | Causes |
|---|---|
| High-pressure switch for refrigeration unit or over load relay protection Circuit works. | Reduce heat load. |
| High-pressure switch for refrigeration unit or over load relay protection Circuit works. | Use the unit where the ambient temperature is lower than 35℃. |
| Refrigeration unit is impaired. | Stop operation and contact your local dealer or closest customer service center. |
| Water level lowers and cooling coil is exposed, which makes the unit perform over load operation. | Fill circulation liquid in low temperature circulation bath. |
| Symptom | Causes |
|---|---|
| Setting temperature is not appropriate. | Check the setting temperature and calibrated displayed temperature. |
| Refrigeration unit does not work. | Stop operation and contact your local dealer or closest customer service center. |
| Fan for refrigeration unit does not work. | Stop operation and contact your local dealer or closest customer service center. |
| Gas is leaking. | Stop operation and contact your local dealer or closest customer service center. |
| Symptom | Causes |
|---|---|
| Gas is leaking. | Stop operation and contact your local dealer or closest customer service center. |
| Fan for refrigeration unit does not work. | Stop operation and contact your local dealer or closest customer service center. |
| Dirt adheres to the filter of refrigeration unit. | Clean the filter. (Refer to the section "Maintenance and checkup" on page 25) |
| Symptom | Causes |
|---|---|
| Valve for adjusting discharge volume is closed. | Open the valve. |
| Dirt adheres to the strainer of low temperature circulation bath. | Remove the dirt. |
| Air is sucked in. | Remove drain stopcock and make sure that cool liquid flows, then attach the stopcock again. Turn on and off pump switch two or three times to release the air. |
| Thermal protector of circulation pump works. | If the cooling liquid is high-viscosity antifreeze, water it down or change the liquid to the one that has low viscosity. |
| Symptom | Causes |
|---|---|
| Replace the circulation nozzle with the other one that has smaller diameter. | Replace with the supplied nozzle that has larger diameter. |
| Hose is crushed. | Stretch the hose. |
| Pressure loss of circulation system is too much. | Reduce the pressure loss. |
| Symptom | Causes |
|---|---|
| High-pressure switch for refrigeration unit or over load relay holding circuit for refrigeration unit works. | ・Set the ambient temperature at 35℃ or lower. ・If the heat load is out of cooling capability, reduce the load. ・Check the fan for refrigeration unit. ・Check the power and voltage. Please make the rating voltage 220V±10% |
| Symptom | Causes |
|---|---|
| Temperature sensor circuit is disconnected. | Stop operation and contact your local dealer or closest customer service center. |
| Symptom | Causes |
|---|---|
| Measured temperature exceeds the upper limit. | After confirming that the bath temperature falls below the measurable range, press the [Set] key to clear the alarm. *Measurable temperature range: -50℃~ *80℃ |
| Symptom | Causes |
|---|---|
| Measured temperature exceeds the lower limit. | After confirming that the bath temperature rises above the measurable range, press the [Set] key to clear the alarm. *Measurable temperature range: -50℃~ *80℃ |
| Symptom | Causes |
|---|---|
| ・Temperature controller is in abnormal status because of noise and etc. ・Ambient temperature exceeds 35℃. |
・Change the installation site. ・Set the room temperature at 35℃ or lower. If that failure mode still remains, Stop operation and contact your local dealer or closest customer service center. |
| Symptom | Causes |
|---|---|
| Poor circulation and bath is not stirred sufficiently because of pressure loss of piping and etc. | ・Open the valve for adjusting discharge volume ・Use antifreeze. ・Using high-pressure pump is recommended. |
| Temperature controller is broken or refrigeration unit cannot be stopped. | Stop operation and contact your local dealer or closest customer service center. |