Trouble shooting
HOME > Trouble shooting > Constant Temperature Bath > CCA-1111-CE
EYELA as a pioneer of Evaporators
HOME > Trouble shooting > Constant Temperature Bath > CCA-1111-CE
Symptom | Causes |
---|---|
Power plug is disconnected to outlet, or not connected to outlet completely. | Turn off residual current device and power switch, and connect power plug to outlet. |
Power source is not supplied. | Turn on the breaker on distribution board. |
Fuse has blown. (Excess current flows.) |
Replace with new fuse. If that failure mode still remains after replacing the fuse, stop running the product immediately and contact your local dealer or closest customer service center. |
Power switch is impaired. | Stop running the product immediately and contact your local dealer or closest customer service center. |
Temperature controller is impaired. | Stop running the product immediately and contact your local dealer or closest customer service center. |
Symptom | Causes |
---|---|
Refrigeration unit is impaired. | Stop running the product immediately and contact your local dealer or closest customer service center. |
SSR is impaired. | Stop running the product immediately and contact your local dealer or closest customer service center. |
Water level lowers and cooling coil is exposed, which causes over load operation. | Refill the cool water. (Refer to「3.Filling cool water」on page 16.) |
Protective timer for refrigeration unit works. | After passing the time, make sure that refrigeration unit works. |
Symptom | Causes |
---|---|
Refrigeration unit does not work. | Stop running the product immediately and contact your local dealer or closest customer service center. |
Fan of refrigeration unit does not work. | Stop running the product immediately and contact your local dealer or closest customer service center. |
Gas leaks. | Stop running the product immediately and contact your local dealer or closest customer service center. |
Symptom | Causes |
---|---|
Gas leaks. | Stop running the product immediately and contact your local dealer or closest customer service center. |
Fan for refrigeration unit does not work. | Stop running the product immediately and contact your local dealer or closest customer service center. |
Ambient temperature is higher than 35℃. | Set the temperature lower than 35℃. |
Heat load is too heavy. | Reduce the load to the value that is within specified range. |
Obstacle closer to the main units prevents emission and exhaust heat. | Set enough space around the unit. (Refer to「4-2. Condition 」on page 12. |
Symptom | Causes |
---|---|
Strainer in the cool water circulation bath has some dust. | Remove the dust. |
Air is sucked in. | Remove drainage plug and make sure that cool water flows from drainage port, and put drainage plug again. Turn on and off the circulation pump for 2-3- times repeatedly to release the air. (Refer to「2. Starting operation」on page 18. |
Stop valve is closed. (In case that any option is attached.) |
Open the valve. (In case that any option is attached.) |
Symptom | Causes |
---|---|
Hose is crushed. | Fix the hose. |
Pressure loss of circulation system is too big. | Reduce pressure loss. |
External circulating point is too high. | Adjust the position. |
Symptom | Causes |
---|---|
Due to pressure loss of pipe, circulation volume is small and bath can not be stirred sufficiently. | Open stop valve. Use antifreeze. |
Temperature controller is impaired or refrigeration unit does not stop. | Stop running the product immediately and contact your local dealer or closest customer service center. |
Symptom | Causes |
---|---|
Ambient temperature exceeds 35℃. | Set the temperature lower than 35℃. |
Heat load is out of the capacity of cooling capability. | Set the heat medium within specified range. |
Dirt adheres to air filter. | Clean the filter. |
Fan in the refrigeration unit does not work. | Check the status of the fan. |
Voltage is too low. For voltage dipped. |
Check the voltage. |
Refrigeration unit is impaired. | 1Reboot the unit adter about 10 minutes. |
Symptom | Causes |
---|---|
Power failure occurred during control. | Make sure to turn off the power switch after stopping control. Power failure alarm can be released by 「Set/Clear」key. *Power failure alarm is availale to select whether power failure alarm is activated or not.(Refer to P.28「How to set power failure alarm」) |
Power Switch is turned off without stopping control. | Make sure to turn off the power switch after stopping control. Power failure alarm can be released by 「Set/Clear」key. *Power failure alarm is availale to select whether power failure alarm is activated or not.(Refer to P.28「How to set power failure alarm」) |
For voltage dipped. | Check the voltage. |
Symptom | Causes |
---|---|
Sensor is disconnected. | Stop running the product immediately and contact your local dealer or closest customer service center. |
Symptom | Causes |
---|---|
Poor stirring in the bath. | Refer to previous page and check the trouble, cause of the trouble and countermeasures for sections related to circulation system. |
Refrigeration units is cooled down poorly. | Refer to previous page and check the trouble, cause of the trouble and countermeasures for sections related to refrigeration unit. |
Symptom | Causes |
---|---|
Poor stirring in the bath. | Refer to previous page and check the trouble, cause of the trouble and countermeasures for sections related to circulation system. |
Symptom | Causes |
---|---|
Due of to power source noise and etc, watch dog of controlling basal plate worked. | Change the power source and reboot the unit. If that failure mode still remains, stop running the product immediately and contact your local dealer or closest customer service center. |
Ambient temperature exceeds 40℃. | Set the temperature lower than 40℃. If that failure mode still remains, stop running the product immediately and contact your local dealer or closest customer service center. |